Measuring device with digital display for monitoring conductivity. A limit value can be set using the rotary switch. The current conductivity can be checked on the LCD display.
Display | digital 3.5-digit LCD display |
---|---|
Measuring ranges | 0-1,999μS/cm 0-19,99μS/cm 0-199,9μS/cm 0-1999μS/cm |
Alarm value display | Limit values adjustable via rotary code switch (10 switching points freely selectable) |
Installation | Wall mounting or for screwing onto a cartridge |
Measuring cell | screwed or measuring cell with 2m cable and plug |
Measuring cell factor | C=0,6 or C=0,1 |
Optional switching contact | Connection of solenoid valve switching contact 24VDC, executed with solenoid valve plug or designed as a potential-free changeover contact, loadable up to 2A/250AC |
Temperature compensation | optional |
Power connection | 9V plug-in power supply 100-240VAC/9DC for solenoid valve: power supply 100-240VAC/24VDC |
Housing material | ABS |
Dimensions in mm | 125 x 70 x 35 |
Front foil / company logo | Layout as per customer request |